Pressure measuring transducer for differential pressure and volume flow
PREMASGARD® 612x - EtherCAT P
The series of networkable PREMASGARD® 612x - EtherCAT P pressure transmitters with M8 plug-in connector (EtherCATP-encoded) are Bluetooth-enabled and have an impact resistant plastic housing with quick-locking screws, optionally with/without display and bar graph, and with nozzles for connecting a pressure hose (Ø 6 mm). The are supplied including an ASD-06 connection kit (2m connecting hose, two pressure port nipples, screws).
The on-wall sensor devices serve to measure positive, negative or differential pressure in clean air and gaseous media as well to calculate volume flows using the K-factor with function selection. The piezoresistive measuring element is temperature-compensated and guarantees a high degree of reliability and accuracy. The sensor is factory-calibrated.
The pressure transducers are used in cleanroom, medical and filter engineering, in ventilation and air conditioning ducts, in spray booths and large-scale catering facilities as well as in filter monitoring and level measuring or frequency converter triggering. In cases of direct solar irradiation, we recommend using our WS 03 weather and sun protection hood (accessory).
The on-wall sensor devices serve to measure positive, negative or differential pressure in clean air and gaseous media as well to calculate volume flows using the K-factor with function selection. The piezoresistive measuring element is temperature-compensated and guarantees a high degree of reliability and accuracy. The sensor is factory-calibrated.
The pressure transducers are used in cleanroom, medical and filter engineering, in ventilation and air conditioning ducts, in spray booths and large-scale catering facilities as well as in filter monitoring and level measuring or frequency converter triggering. In cases of direct solar irradiation, we recommend using our WS 03 weather and sun protection hood (accessory).
EtherCATP-enabled measuring transducer for industrial requirements with maximum reliability: includes easy PLC integration using the devices ESI configuration file, diagnostics (such as communication failure counter), advanced setting options, access to historical data (min/max) and sensor service interval determination. Optionally with large back-lit three-line display (customised programming in the 7-segment and dot-matrix ranges) and bar graph (7-digit, LEDs freely configurable) for graphical display, e.g. as a traffic light indicator.
Technical details
Voltage supply
24VDC via EtherCATP (US)
Power consumption
< 3 W
Measuring range, pressure
– 500...+ 500 Pa
Radio technology
Bluetooth (LE)
Media temperature
–20...+50 °C (temperature-compensated 0...+50 °C)
Pressure connection
with connection nozzles for pressure hose Ø 6 mm
Type of pressure
Differential pressure [Pa],
volume flow [m3/h]
volume flow [m3/h]
Above- below-pressure
máx. ± 50 kPa
Medium
clean air and
non-aggressive, non-combustible gases
non-aggressive, non-combustible gases
Cable connection
M8 plug-in connector, EtherCATP-encoded
Dimensions
126 x 90 x 50 mm (Tyr 2)
Pressure range
±500Pa
Long-term stability
± 1 % per year
Hysteresis
0.3 % of final value
Media contacting parts
Brass, Ni, Duroplast, Si, epoxy, RTV, BSG, UV silicone gel
Temperature drift values
± 0.1 %/ °C
Linearity
< ± 1 % of final value
Permissible air humidity
< 95 % RH, non-precipitating air
Protection class
III (according to EN 60 730)
Protection type
IP 54 (according to EN 60 529) when built-in
Standards
CE conformity, according to EMC Directive 2014/ 30/ EU,
according to RED 2014/ 53/ EU
according to RED 2014/ 53/ EU
Accuracy
± 3 Pa
Accuracy, pressure
typically ± 3 Pa at +25 °C
compared to the calibrated reference device
compared to the calibrated reference device
Measuring range pressure
– 500...+ 500 Pa
Housing
plastic, UV-stabilised,
material polyamide, 30 % glass-globe reinforced,
with quick-locking screws
(slotted/ Phillips head - combination),
colour traffic white (similar to RAL 9016),
housing cover for display is transparent !
material polyamide, 30 % glass-globe reinforced,
with quick-locking screws
(slotted/ Phillips head - combination),
colour traffic white (similar to RAL 9016),
housing cover for display is transparent !
Features
Display with illumination, 3-line, customised programming,
cutout approx. W51 x H29mm, to display the actual temperature
or a customised display value.
Bar graph, 7-digit, LEDs freely configurable,
to graphically display the reading.
cutout approx. W51 x H29mm, to display the actual temperature
or a customised display value.
Bar graph, 7-digit, LEDs freely configurable,
to graphically display the reading.