Pressure and differential pressure measuring transducer PREMASGARD 1211-M

Pressure and differential pressure measuring transducer PREMASGARD 1211-M

Item no. 1301-1214-0010-200
EAN 4251325612188
Product group: WG01

In Stock

Usually ships in 1 to 3 days
161,80 €
192,54 € incl. VAT + shipping
+

Description

Pressure and differential pressure measuring transducer PREMASGARD® 121x - Modbus
The calibratable, compact pressure measuring transducers of the PREMASGARD® 121x - Modbus series with Modbus connection can be equipped with an optional display and are used for measuring above-atmospheric, below-atmospheric or differential pressures in the air. The piezoresistive measuring element is temperature-compensated and guarantees a high degree of reliability and accuracy.
The pressure transmitters feature a configurable offset and a button for manual zero adjustment. Applications of these sensor devices are found in cleanroom, medical and filter engineering, in ventilation and air conditioning ducts, in spray booths and large-scale catering facilities, in filter monitoring and level measuring or frequency converter triggering. Media measured with these pressure transducers are air (non-precipitating) or other gaseous, non-aggressive, non-combustible media. The devices are supplied including an ASD-06 connection set (2 m connection hose, two pressure connection nipples, screws).

Technical details

Voltage supply
24 V AC (± 20 %) and
15…36 V DC
Power consumption
< 1,0 W/ 24 V DC
< 2,2 VA/ 24 V AC
Pressure range
±1000Pa
Measuring ranges
–1000...+1000 Pa or
–5000...+5000 Pa,
depending on the type of device
Media temperature
–20...+50 °C
Pressure connection
4/ 6 x 11 mm (hoses Ø = 4/ 6 mm),
metal pressure connection nozzles
Type of pressure
differential pressure
Accuracy
(1000 Pa): typically ± 5 Pa
compared to the calibrated reference device
Accuracy, pressure
± 5 Pa
Zero point offset
± 10 % of pressure range
(selectable using the offset calibration button with potentiometer on
circuit board or using the Modbus interface by remote maintenance)
Above- below-pressure
max. 5 x pressure range
Medium
clean air and
non-aggressive, non-combustible gases
Housing
plastic, UV-stabilised,
material polyamide, 30 % glass-globe reinforced,
with quick-locking screws
(slotted/ Phillips head - combination),
colour traffic white (similar to RAL 9016)
Housing dimensions
72 x 64 x 37,8 mm (Tyr 1)
Cable gland
2x M 12 x 1.5 (Y-adapter); with strain relief,
exchangeable, max. inner diameter 6 mm
Long-term stability
± 1 % per year
Hysteresis
0.3 % of final value
Media contacting parts
ms, Ni, Nylon, PU, Si, PVC with plasticisers
Temperature drift values
± 0.1 % of final value/ °C
Linearity
< ± 1 % of final value
Permissible air humidity
< 95 % RH, non-precipitating air
Protection class
III (according to EN 60 730)
Protection type
IP 65 (according to EN 60 529)*
Standards
CE conformity,
electromagnetic compatibility according to EN 61326,
according to EMC Directive 2014/ 30/ EU
MODBUS-Y
Y-adapter for cable gland M16 x1.5 (on 2x M12 x1.5),
plastic (included in the scope of delivery)

Accessories